See how objective NA and aperture affect the thickness that appears in focus.
Choose your objective
Colour bands are a general guide; use the NA marked on your own objective.
Match the two markings on your eyepiece, for example 10×/22. Eyepiece magnification affects the visual-focus estimate; FN changes the visible specimen width, not the calculated depth. Camera viewing needs a different criterion.
For fluorescence, use the detected emission wavelength. White light uses 550 nm as an example.
Objective aperture
Approximate diffraction depth of field
1.30 µm
Physical depth of focus · specimen side0.65 µmFull-aperture diffraction component
Visual depth of focus · through eyepieces1.20 µmIncludes an eyepiece viewing term; 10× eyepieces selected.
Approximate visible field diameter · specimen side0.55 mmFN 22 ÷ 40× objective. Assumes 1× tube factor and sufficient objective field.
Depth of field at full objective NA1.30 µm
Effective objective NA0.65
Side view: objective, aperture and specimen
40× / 0.65 NA · Biological dry
Specimen layersDepth appearing in focus
The blue cone shows collected light angle. The highlighted specimen thickness grows as the simulated objective iris closes. Schematic, not to scale.
General guide only. Values are theoretical approximations, not measured specifications or guaranteed performance. Illumination-aperture slider values are illustrative. Actual focus range varies with objective design, the glass or plastic thickness beneath or above the specimen, sample depth and refractive index, immersion medium, wavelength, contrast method, illumination and microscope adjustment. Check the objective documentation and verify on your own system for critical work.
Compare two objectives at full aperture
The same selected wavelength is used for both. Each lens shows its full-aperture depth of field and physical focus-depth convention in microns.
Your selected objective40× / 0.65 NA1.30 µmPhysical focus depth: 0.65 µm Through eyepieces: 1.20 µm
8.80 µmPhysical focus depth: 4.40 µm Through eyepieces: 10.1 µm
Bar lengths compare the diffraction term on a logarithmic scale, so very low and high NA objectives remain visible. Micron values are the useful comparison.
What changes when you adjust an aperture?
Depth of field · specimen side
Thickness of the specimen that looks acceptably sharp. The full-aperture estimate above is in microns. Closing an adjustable objective iris increases the calculated estimate.
Illumination diaphragms
A transmitted condenser aperture or reflected-light incident aperture can change apparent depth and contrast. The slider illustrates the change in apparent focus, while the listed micron values remain full-objective-NA baselines. Its percentage is not a calibrated NA or micron scale.
Physical and visual focus depth
Your table’s physical value is λn/(2NA²): 50× / 0.80 NA at 550 nm gives 0.43 µm. The visual eyepiece estimate adds the glossary’s magnification term. Both describe specimen-side focus ranges; sensor-side tolerance is different.